Metrology.NET®

TestProcess.Source.Ratio.Voltage.AC.Sinewave.Delta.Frequency

A test process the sources 2 sinusoidal AC Voltage signals at different frequencies keeping the voltage levels constant between the two frequencies. The FrequencyRef signal is sourced and the voltage is measured by the UUT and saved as the Reference value.  Then the signal is changed to the test Frequency signal and the Test value is measured on the UUT.  The two measured values are compared, the results can be expressed as Delta Volts, dB or percent of change.

These two sinusoidal signals are often used when calibrating bandwidth or frequency response on an Oscilloscope or similar device.

Required Parameters

  • Volts – RMS
  • FrequecyRef –
  • Frequency –

Optional Parameters

  • Impedance –
  • UUT.Input – Input Name of the UUT Connecting Point

Output Value & Uncertainty

  • dB – 20 * Log( Measured Value / Reference Value)
  • % – Measured Value / Reference Value * 100

Leave a Reply