The process of sourcing 3 sinusoidal AC Voltage signals from a device with and measuring the value on the UUT.
This test process can be used by any device that measures absolute AC Voltage on 3 separate legs.
Required Parameters
- Volts – RMS
- Frequency – Hz
Optional Parameters
- Phase – 120 deg
- Impedance –
- UUT.Input – Input Name of the UUT Connecting Point
Output Value & Uncertainty
The process of sourcing 2 sinusoidal AC Voltage signals from a device with and measuring the value on the UUT.
This test process can be used by any device that measures absolute AC Voltage on 2 separate legs.
Required Parameters
- Volts – RMS
- Frequency – Hz
Optional Parameters
- Phase –
- Impedance –
- UUT.Input – Input Name of the UUT Connecting Point
Output Value & Uncertainty
This test process generates the pneumatic absolute pressure to be measured by the unit under test. The resulting measured will be in the unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – {μmHG, mmHG, mHG, μmH2O, mmH2O, mH2O, inHg, inH2O, psi, μPa, mPa, Pa, kPa, MPa, mbar, bar, Torr}
Optional Parameters
- O2_Safe – “Yes” If the Gage in an O2 environment
Measured Value & Uncertainty
Measured Value & Uncertainty
Pressure
This test process generates the hydraulic pressure to be measured by the unit under test. The resulting measured will be in the unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – {μmHG, mmHG, mHG, μmH2O, mmH2O, mH2O, inHg, inH2O, psi, μPa, mPa, Pa, kPa, MPa, mbar, bar, Torr}
Optional Parameters
Measured Value & Uncertainty
This test process generates two pneumatic pressures with a known differential pressure between two points measured by the unit under test. The resulting differential pressure will return in the unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – {μmHG, mmHG, mHG, μmH2O, mmH2O, mH2O, inHg, inH2O, psi, μPa, mPa, Pa, kPa, MPa, mbar, bar, Torr}
Optional Parameters
- O2_Safe – “Yes” If the Gage in an O2 environment
Measured Value & Uncertainty
This test process generates the pneumatic gage pressure (or relative pressure) to be measured by the unit under test. The resulting measured will be unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – {μmHG, mmHG, mHG, μmH2O, mmH2O, mH2O, inHg, inH2O, psi, μPa, mPa, Pa, kPa, MPa, mbar, bar, Torr}
Optional Parameters
- O2_Safe – “Yes” If the Gage in an O2 environment
Measured Value & Uncertainty
This test process measures the hydraulic pressure generated by the unit under test. The resulting measured will be in the unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – { psi, bar}
Optional Parameters
Measured Value & Uncertainty
This test process measures the pneumatic differential pressure between two points generated by the unit under test. The resulting pressure difference will return in the unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – {μmHG, mmHG, mHG, μmH2O, mmH2O, mH2O, inHg, inH2O, psi, μPa, mPa, Pa, kPa, MPa, mbar, bar, Torr}
Optional Parameters
- O2_Safe – “Yes” If the Gage in an O2 environment
Measured Value & Uncertainty
This test process measures the pneumatic gage (or relative pressure) generated by the unit under test. The resulting measured will be in the unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – {μmHG, mmHG, mHG, μmH2O, mmH2O, mH2O, inHg, inH2O, psi, μPa, mPa, Pa, kPa, MPa, mbar, bar, Torr}
Optional Parameters
- O2_Safe – “Yes” If the Gage in an O2 environment
Measured Value & Uncertainty
This test process measures the pneumatic absolute pressure generated by the unit under test. The resulting measured will be in the unit of measure as defined in the test point’s data.
Note: Although the unit of measure can be read from the test point’s unit of measure it is good practice to pass the Unit of Measure to the test process as a parameter to avoid confusion when synchronizing data between systems.
https://en.wikipedia.org/wiki/Pressure_measurement
https://www.nxp.com/docs/en/application-note/AN1573.pdf?&srch=1
Required Parameters
- Pressure
- UnitOfMeasure – {μmHg, mmHg, mHg, μmH2O, mmH2O, mH2O, inHg, inH2O, psi, μPa, mPa, Pa, kPa, MPa, mbar, bar, Torr}
Optional Parameters
- O2_Safe – “Yes” If the Gage in an O2 environment
Measured Value & Uncertainty